Fluke 9132/9133 Portable Infrared Calibrator

Precision when you need it for infrared temperature calibration

  • Certify IR pyrometers from Ð30 ¡C to 500 ¡C (Ð22 ¡F to 932 ¡F)
  • Large 57 mm (2.25 in) blackbody target
  • RTD reference well for contact temperature measurement
  • Small, compact design

Whether youÕre using in-line or handheld infrared pyrometers, you need good calibration standards to verify their accuracy. Our portable IR calibrators provide stable blackbody targets for calibrating noncontact IR thermometers from Ð30 ¡C to 500 ¡C.

These units feature a temperature controlled measurement surface with a diameter of 2.25" (57 mm). The target temperature can be controlled in set-point increments of 0.1 ¡ from Ð30 ¡C to 500 ¡C and a well is located directly behind the blackbody surface for contact calibration.



Key features

9132

For IR calibrations above normal ambient, the 9132 provides a stable measurement surface up to 500 ¡C (932 ¡F). Short heating and cooling times mean you won't have to wait long to get your work done.

9133

If you're calibrating IR guns at cold temperatures, you'll love our new 9133. With solid-state cooling technology, this new IR calibrator reaches Ð30 ¡C (22 ¡F) in normal ambient conditions. With a conveniently located dry gas fitting on the front bezel, ice build up on the target can be avoided. At the upper end of its range, the 9133 provides stable temperatures to 160 ¡C (320 ¡F).

With heating and cooling times of about 15 minutes from ambient to either extreme, the 9133 gets you to temperature quickly and performs when it gets there.



Specifications91329133
Temperature range50¡C to 500¡C
(122¡F to 932¡F)
Ð30¡C to 150¡C at 23¡C ambient
(Ð22¡F to 302¡F at 73¡F ambient)
Accuracy±0.5¡C at 100¡C
(±0.9¡F at 212¡F)
±0.8¡C at 500¡C
(±1.4¡F at 932¡F)
±0.4¡C (±0.72¡F)
Stability±0.1¡C at 100¡C
(±0.18¡F at 212¡F)
±0.3¡C at 500¡C
(±0.54¡F at 932¡F)
±0.1¡C (±0.18¡F)
Target size57 mm (2.25 in)57 mm (2.25 in)
Target emissivity0.95 (±0.02 from 8 to 14 mm)0.95 (±0.02 from 8 to 14 mm)
Resolution0.1¡0.1¡
Heating time30 minutes (50¡C to 500¡C)15 minutes (25¡C to 150¡C)
Cooling time30 minutes (500¡C to 100¡C)15 minutes (25¡C to Ð20¡C)
Computer interfaceRS-232 I/O included
Power115 V AC (±10%), 3 A or
230 V AC (±10%), 1.5 A,
switchable, 50/60 Hz, 340 W
115 V AC (±10%), 1.5 A, or
230 V AC (±10%), 1.0 A,
switchable, 50/60 Hz, 200 W
Size (H x W x D)102 x 152 x 178 mm
(4 x 6 x 7 in )
152 x 286 x 267 mm
(6 x 11.25 x 10.5 in)
Weight1.8 kg (4 lb)4.6 kg (10 lb)
NIST-traceable contact calibrationData at 50¡C, 100¡C, 200¡C, 250¡C, 300¡C, 400¡C, and 500¡CData at Ð30¡C, 0¡C, 25¡C, 75¡C, 100¡C, 125¡C, and 150¡C