Fluke Calibration P3200 Hydraulic Deadweight Tester

The P3200 models are available in single or dual piston formats for increased operating ranges. Units can be supplied in psi, bar, kgf/cm_, and MPa. This robust instrument is highly accurate, quick and easy to use. Units feature a built-in priming pump for large volume applications, piston flotation indicators and a high quality screw press for fine pressure control.

Options

PressCal software for pressure calculations and certificate generation



Key features

  • Pressure ranges to 10,000 psi (700 bar)
  • 0.015% of reading accuracy standard (0.008% optional)
  • Single or dual piston formats
  • Built-in hand pumps standard
  • Mounted spirit level with adjustable feet
  • Units can be trimmed to local gravity FOC


Pressure Ranges
Distilled water operatedRanges to 10,000 psi (700 bar)
Accuracy
Accuracy1±0.015% of reading (±0.008% optional)
1. Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range.
Materials of Construction
Standard weight materialSeries 3 non-magnetic austenitic stainless steel
Weight density7.8 g/cm_
Optional fractional weightsSolution heat treated aluminum
Weight density2.7 g/cm_
Piston materialTungsten carbide with nickel binder
Density- 15.0 g/cm_
Cylinder materialHardened martensitic steel
Tungsten carbide (water above 500 psi, 35 bar)
Thermal coefficients of expansionOil piston/cylinder 16.5 ppm/¡C
Water (above 500 psi, 35 bar) 11 ppm/¡C
General
Test port adaptors1/8, 1/4, 3/8 and 1/2
NPT or BSP
Instrument weight36 lbs (16 kg)
Instrument size (W x D x H)17.5 x 12 x 8.5 in (440 x 300 x 215 mm)
Mass set weight (typical)80 lbs (36 kg)
Reservoir volume150 cc
Screw press displacement5.5 cc
Pump displacement4.7 cc per stroke
O-ring seal materialsBuna N as standard, Viton and EPDM available
Weight Increments
Minimum standard weight incrementsUp to 500 psi (35 bar): 1 psi (0.1 bar)
500 to 5000 psi (35 to 350 bar): 10 psi (1 bar)
5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar)
Model P3112: 1 psi (0.1 bar)
Optional fractional weightsUp to 500 psi (35 bar): 0.1 psi (0.01 bar)
500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar)
5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar)
Operating Fluids
Operating fluidsDistilled or deionized water
Options
PressCal softwareWindows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments
FluidsInstruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory.